V.A.C.Via™ Therapy System Product Specifications*
Feature
Description
Unit Weight with empty canister
0.7 lbs (0.32 kg)
Dimensions with empty canister
Length 6.8" (17.2 cm); Width 3.9" (9.9 cm); Depth 2.4" (6.1 cm)
Pump
Two piezoelectric disc pumps
User Interface
Membrane switch; 3-second press and hold
Therapy Delivery Modes
Continuous or Dynamic Pressure Control™
(DPC Therapy)
(DPC Therapy)
Therapy Pressure Options
Negative pressure: -75 mmHg or -125 mmHg
Power Source
AC and battery
Battery Type
Custom lithium battery
Battery Life
~ 9 hours
Battery Recharge Time
Less than 6 hours
Battery Indicator
Three-segment; low-battery alarm when
<2 hours
<2 hours
Canister
250mL, omni-directional, with Isolyzer
Safety Alarms
System Fault, Blockage, Leakage, Battery Critical, Therapy Life Exhausted
Electrical Data
External power supply input: 100-240
V.A.C. @ 50v60Hz
External power supply output: nominally 9Vdc (18 Watt)
Patient & Endosure Leakage Current: <100μA @ 132 Vac @ 60 Hz
V.A.C. @ 50v60Hz
External power supply output: nominally 9Vdc (18 Watt)
Patient & Endosure Leakage Current: <100μA @ 132 Vac @ 60 Hz
Electromagnetic Compatibility
Electromagnetic interference: Although this equipment conforms with the intent of the directive 89/336/EEC in relation to Electromagnetic Compatibility (EMC), all electrical equipment may produce interference. If interference is suspected, move equipment away from sensitive devices or contact the manufacturer. Radios, cell phones and similar devices may affect this equipment and should be kept at least 6.5 feet (2 meters) away from the equipment.
IEC Classification
EC 60501-1 Classification:
Medical Electrical Equipment
Type BF applied part
Class II equipment
IP22 ingress protection
Medical Electrical Equipment
Type BF applied part
Class II equipment
IP22 ingress protection
Environmental Conditions
Transport and Storage Conditions:
Temperature Range: -20°F (5°C) to 140°F (60°C)
Relative Humidity Range: 0-95%, non-condensing
Temperature Range: -20°F (5°C) to 140°F (60°C)
Relative Humidity Range: 0-95%, non-condensing
Operating Conditions
Temperature: 41°F (5°C) to 104°F (40°C)
Relative Humidity: 15-95%, non-condensing
Altitude Range: Optimal Performance: 0 to 8,000 feet (0 to 2438m)
Relative Humidity: 15-95%, non-condensing
Altitude Range: Optimal Performance: 0 to 8,000 feet (0 to 2438m)
Packaging Dimensions
V.A.C.Via™ System Starter Kit Shipper: 44 x 25.6 x 58.4 cm
V.A.C.Via™ System Starter Kit Each: 42.5 x 24.7 x 10.6 cm
V.A.C.Via™ System Medium Dressing Shipper: 31.5 x 19 x 26.4 cm
V.A.C.Via™ System Medium Dressing Pouch: 31.5 x 25.1 cm
V.A.C.Via™ System Small Dressing Shipper: 31.4 x 19 x 26.4 cm
V.A.C.Via™ System Small Dressing Pouch: 31.5 x 19 x 26.4 cm
V.A.C.Via™ System Canister Shipper: 46 x 20.8 x 14.6 cm
V.A.C.Via™ System Starter Kit Each: 42.5 x 24.7 x 10.6 cm
V.A.C.Via™ System Medium Dressing Shipper: 31.5 x 19 x 26.4 cm
V.A.C.Via™ System Medium Dressing Pouch: 31.5 x 25.1 cm
V.A.C.Via™ System Small Dressing Shipper: 31.4 x 19 x 26.4 cm
V.A.C.Via™ System Small Dressing Pouch: 31.5 x 19 x 26.4 cm
V.A.C.Via™ System Canister Shipper: 46 x 20.8 x 14.6 cm
*Specifications subject to change without notice.
NOTE: Specific indications, contraindications, warnings, precautions and safety information exist for KCI products and therapies. Please consult a physician and product instructions for the price to application. CAUTION: Federal law restricts this device to sale by or on the order of a physician.
For more information on V.A.C.® Therapy and other KCI products, click here.